AMPROBE
DMIII MULTITEST
EN-35
THE PREVIOUS RESULTS CANNOT BE SAVED.
If the Instrument
detects a Resistance
value higher than
1999, the instrument
will show the screen to
the right.
EARTH 05.06.01
Message ">1999" means that
the resistance value is higher
than the maximum
measurable.
> 1999
Vd= 1V
Test:04
RAVG=0.74
3-W
FUNC
CLR
This result can be stored pressing the SAVE key twice (refer to paragraph 9.1).
If the Instrument
detects a Resistivity
value higher than
1999km, the
instrument will show
the screen to the right.
EARTH 05.06.01
Message ">1999" means that
the resistivity value is higher
than the maximum
measurable.
>
1999km
Vd= 1V
Test:04
AVG=0.74km
DIST=5m
FUNC
CLR
This result can be stored pressing the SAVE key twice (refer to paragraph 9.1).