Troubleshooting
Problem Solution
Thermometer does not turn
on.
Check battery voltage and
replace discharged battery.
Verify that battery clips grip
the battery post tightly.
Dashes appear in the T1 and
T2 data screens.
Insert missing
thermocouple.
Dashes appear in T1 and/or T2
data screens with
thermocouples inserted.
Measure resistance of
thermocouples to ensure
they are not broken
internally.
Clean the thermocouple
and restart.
Temperature drifts from a
known value in a controlled
environment
Verify that thermocouple
type matches the displayed
icon.
Clean and dry the
thermocouple blades and
allow to air dry.
Confirm temperature with a
known good thermocouple.
Replace the defective
thermocouple if required.
TMD90A_all.pdf 12 3/23/2007 3:22:49 PM