5
SYMBOLS
Refer to the manual
Complies with EU
directives
Conforms to relevant
Australian standards.
Do not dispose of this
product as unsorted
municipal waste.
Warning and Precautions
When using Sound-Level Calibrator, ambient noise level should > 10dB
below the sound level being measured.
UNPACKING AND INSPECTION
Your shipping carton should include:
1 SM-CAL1 sound generator
1 1/2” adaptor
1 9 volt battery
1 Manual
If any of the items are damaged or missing, return the complete package to
the place of purchase for an exchange.
INTRODUCTION
The SM-CAL1 generates a fixed sound level signals for calibration of sound
level meters. The Amprobe SM-CAL1 will work with any sound meter.
OPERATION
Turn on sound level meter and configure for 94 dB or 114 dB operation.
Plug the 1/2 adapter into the SM-CAL1 Sound level calibrator.
Insert any sound meter microphone into the Sound-Level Calibrator.
Turn on power switch and set sound level to 94 dB or 114 dB.
Observe the sound level meter for proper operation.
•
1.
2.
3.
4.
5.
Komentarze do niniejszej Instrukcji