RANG Resolution Accuracy *V Maximum
-40°C ~1372°C 1°C
-40°F ~ 2502°F 1°F
±(0,5%rdg+3°C)
±(0,5%rdg+6°F)
30 V RMS
or 60 V DC
c
K-TYPETEMPERATUREMEASUREMENTRANGE
Gama Resolución Precisión *V Máximo
-50°C< T<100°C 1°C
-58°F< T<180°F 1°F
±(0,5%rdg+3°F)
±(0,5%rdg+6°F)
30 V RMS
or 60 V DC
c
TEMPERATURADIFERENCIALTIPO K
RANGE Resolution Accuracy Overload Protection
400µF 0.1µF
400µF 1µF
±(3%rdg+4dgt)
±(3.5%rdg+4dgt),
>2mF, NO Spec
600 V
RMS
c
CAPACITANCE
Note: Do not alow the temperature sensor to contact a surfece which is energized above
30 V RMS or 60 V DC, such voltages pose a shock hazard.
The T1 input must be greater than -T2 input for T1-T2 measuring.
the accuracy does not include the tolerance of thermocouple probe.
MODELTPK-56, BEADTYPE-K THERMOCOUPLE (optional)
Temperature Rating (wire): continuous 204°C, 260°C (intermittent)
Temperature measuring Range (bead): -50
o
C to +800°C (MAX) Accuracy.
±2.2°C or ±0.75% of reading from 0°C to 800°C (whichever is greater)
±2.2°C or ±2% of reading from 0°C to -50
o
C (whichever is greater)
Caution: The bead end can be touched to surfacve temperatures above 204
0
C
for very short durations. However, wire insulation or wire may be damage if wire
subjected to the same temperature.
With film capacitor or better, use ZERO to cancel residual.
Temperature coefficient: 0.15 x specified accuracy / °C(from 18
o
C or 28
o
C to 50
o
C) or 0.084 x
specified accuracy/
o
F(from 32
o
F to 64.4
o
F or 82.4
o
F to 122
o
F)
50