7
OL Protection: 1500 V dc/1000 V ac rms
Transient Protection: 12 kV impulse (1.2 µS/50 µS)
based on EN 61010-1:2001 impulse requirement
for at CAT IV 1000 V product. This product should
not be used in installations where transients
exceed 12 kV.
DC Current
Ranges: 200 µA, 2 mA, 20 mA, 200 mA, 2 A
Resolution: 0.1 µA
Accuracy: 200 µA to 20 mA: ±(0.5 % rdg+2 dgts),
200 mA to 2 A: ±(1.0 % rdg+2 dgts)
Voltage Burden: 250 mV max.
(1 mV/1 µA on 200 µA Range),
(100 mV/1 mA on 2 mA Range),
(10 mV/1 mA on 20 mA Range),
(1.5 mV/1 mA on 200 mA Range),
(500 mV/1 A on 2 A Range)
OL Protection: A Input (F 2 A/1500 V,
size 8 x 65 mm IR fast blow ceramic)
AC Current (45 Hz to 1 kHz)
Ranges: 200 µA, 2 mA, 20 mA, 200 mA, 2 A
Resolution: 0.1 µA
Accuracy: 200 µA to 20 mA: ±(0.8 % rdg+4 dgts),
200 mA to 2 A: ±(1.2 % rdg+4 dgts)
Voltage Burden: 250 mV max. (1 V on 2 A range)
OL Protection: Same as DC Current.
Resistance
Ranges: 200 , 2 k, 20 k, 200 k, 2 M, 20 M
Resolution: 0.1
Accuracy: 200 ±(0.5 % rdg+4 dgts),
2 k to 200 k: ±(0.3 %rdg+2 dgts),
2 M: ±(1.0 % rdg+4 dgts),
20 M: ±(2.0 %rdg+4 dgts)
OL Protection: 1500 V dc/1000 V ac rms
Open Circuit Voltage, 200 rng: 3.0 V dc typical;
all other rngs 0.3 V dc typical.
Diode/Continuity Test
Test Current: 1m A dc typical
Test/Open Circuit Voltage: 3.0 V dc Typical
Diode Accuracy: ±(1.5 % rdg +2 dgts)
Continuity Audible Threshold: < 50 ± 25
Response Time: <100 msec
OL Protection: 1500 V dc/1000 V ac rms