34XR-A Users Supplement
9/06 1
Change #1
On page 9, under General Specification, change
Operating environment:
From: 0 ºC to 50 ºC at <70% R.H.
To: 0 ºC to 50 ºC at <70% R.H. for all functions
except 10A ranges
10A ranges: 0 ºC to 40 ºC at <70% R.H.
Change #2
On page 9, following Operating environment add:
10A Ranges: 0 ºC to 40 ºC at < 70 % R.H.
On page 10, under AC CURRENT, following Crest
Factor add the following:
Specified From: 10 % to 100% of range
On page 11, under FREQUENCY change:
From: Range: 4K, 40K, 400K 4M, 40MHz
To: Range: 4K, 40K, 400K, 1 MHz
Under Sensitivity, change the following and delete,
4 MHz to 40 MHz: >2 V ac rms, <5 V ac rms
From: 10 Hz 4 MHz: >1.5 V acrms
To: 10 Hz to 1 MHz: >2.5 V ac rms
From: Minimum pulse width: > 25 ns
To: Minimum pulse width: > 1 µs